IVI Orlando PC
Our ServiceBuy/Leasing PC EquipmentRate for leasing, buying and service contractsIVI ComputersIVI Computers salesCalibration TechNetComputer System TroubleshootingIVI TechNetPrimary Reconstruction/Volumetric DataContact InformationSupport Question and Trouble TicketsIVI Connect
Calibration TechNet

 Calibration
In order to perform the calibration, the service level must be enable (see chapter 3.3 ” Enabling the service
level”). For machine with multiple FOV the complete calibration of the system consists of two or three
different calibrations, one for each Field Of View (FOV). Each single calibration includes the grid, the
mattarellum, the background and the phantom acquisition and analysis.
NOTE (Only for machine with the multiple FOV):
• Perform the calibration of the device following this sequence: Large, Medium and Small FOV.
• Terminate the calibration of a single FOV before proceeding with the next one.
• At least 10 Gbyte of free space are required in order to perform the complete calibration.
Diaphragm calibration
This procedure is not required in standard calibration, but only in particular cases or after changing of the
image chain.
Software configuration
The software configuration must be done the first time we connect a PC to the machine. It stores a flag that
permit to the software to use the correct parameters to calibrate the machine.
1. Press ALT+F10 and insert the password: 336699
2. In the “General Settings” window, select the “Configuration” tab
3. Set the “Device Type” parameters as follow: “CSA Shell” for device labelled as CSA compliant and
“Standard” for all others.
4. Select OK to confirm.
Diphragm calibration
1. Be sure the plastic frontal panel in mounted
2. From the main menu select Tools Scanner tests
3. Move the arm to 45 degrees
4. From the menu select Tools CCD Camera Pad
5. Select the “Calibr” button.
6. If required select which model of Image Intensifier is mounted on the device (12” or 9”)
7. Wait the end of the automatic procedure
8. Take note of the values reported on the final window. Store them in the device’s documentation.
Grid calibration
1) Select Scan Grid Acquisition.
2) Select the FOV (large , medium, small) to be used for calibration.
Blank acquisition
Before starting the grid acquisition, the software will require the execution of the blank acquisition.
3) Slide the tabletop all the way out of the gantry.
4) From the main window of the software select the “Rx_shot” button.
5) After a few seconds the blank image will appear.
Grid Acquisition
Once the blank acquisition has been performed, the software is ready for the grid acquisition.
6) Mount the calibration Grid on the detector (see above picture).
7) Verify that the tabletop is still all the way out of gantry.
8) From the main window of the software select the “Rx_shot” button.
9) The software will acquire an image of the grid.
10) Once the image is displayed select the “Start” button.
11) The acquisition process will start.
12) At the end of the scan the analysis of the scan will automatically start.
13) After a few minutes the report of the analysis will appear on the screen.
14) Select File Close.
15) Remove the grid from the detector.
Mattarellum calibration
1) Select the FOV (large , medium, small) to be used for calibration (must be the same used for the
previous grid calibration).
Blank acquisition
If required by the software perform the blank acquisition as explained in the following steps.
2) Slide the tabletop all the way out of the gantry.
3) From the main window of the software select the “Rx_shot” button.
4) After a few seconds the blank image will appear. Select the “Stop” button.
Mattarellum Acquisition
Once the blank acquisition has been performed, the software is ready for the mattarellum acquisition.
5) Select Scan Mattarellum Acquisition.
6) Insert the mattarellum in the gantry, with the iron rod in the upper side (see above picture).
7) Verify that the tabletop is still all the way out of gantry.
8) From the main window of the software select the “Rx_shot” button.
9) The software will acquire an image of the mattarellum.
10) Once the image is displayed select the “Start” button.
11) The acquisition process will start.
12) At the end of the scan the analysis of the scan will automatically start.
13) After a few minutes the report of the analysis will appear on the screen.
16) Select File Close.
14) Remove the mattarellum from the detector.
Background acquisition
This calibration corrects the dishomogeneity of the detector. It could be performed also by the user, if rings
are reported on axial images.
1) Slide the tabletop all the way out of the gantry.
2) Select the FOV (large , medium, small) to be used for calibration (must be the same used for the
previous grid and mattarellum calibrations).
3) Select Scan Background Acquisition (If required by the software perform a Daily Check before
proceeding).
4) Once checked the scan area is empty, select the “Next” button.
5) From the main window of the software select the “Rx_shot” button.
6) The software will acquire a scout image.
7) Once the image is displayed select the “Next” button.
8) Select the “Start” button.
9) The background acquisition will start. This process consists in six consecutive scans that are
automatically run by the software. At the end of the sixth scan the acquired data will be analyzed.
10) Once the analysis process is terminated select File Close.
Phantom QA
1) Activate the service level (see chapter: 3.3 )
2) Be careful that all corrective factors are setting to 1 before to start this calibration ( see chapter: 11.6 ).
3) Select the FOV (large, medium, small) to be used (must be the same used for the previous grid,
mattarellum and background calibrations).
4) Run a phantom QA scan as described in the User Manual. Stop the analisys (press the Cancel button
on the reconstruction’s progress bar).
5) Calculate the corrective factors for the selected FOV as explained below:
Large FOV: a) DSD/DSCR (see 11.6.1 ” DSD / DSCR calibration”)
b) Autocalibration (see 11.6.4 ” Autocalibration”)
c) Z factor (see 11.6.2 ” Z factor calculation”)
Medium and small FOV: a) XY factor (see 11.6.3”X-Y factor calculation”)
b) Z factor (see 11.6.2 ” Z factor calculation”)
NOTE (Only for machine with the multiple FOV):
After the phantom analysis is terminated and all the values are in the ranges, the calibration
of the current FOV is complete and it is possible to proceed with the calibration of the
following FOV, starting from the grid acquisition (see 11.2 Grid calibration).
Corrective factors calculation
During the manufacturing of the device, various factors are calculated and stored in the software, in order to
be applied to the different FOV acquisitions. These factors can be viewed in the “General Settings” window
under the “Geometric Parameters” tag (see chapter: 3.4 Opening the “General Settings” window).
After the x-ray source or the x-ray detector has been replaced, these factors may need to be recalculated.
Next are the steps to follow to recalculate these factors.
DSD / DSCR calibration
NOTE: The following procedure applies only to the large FOV.
Next are the steps to follow in order to run the DSD / DSCR calibration.
1. Open the RawData of the last QA phantom scan, performed using the Large FOV.
2. Select Primary reconstruction -> DSD / DSCR calibration.
3. Wait for the procedure to terminate.
4. The software will show the two calculated values and will automatically store them.
5. Perform a new phantom QA scan (using Large FOV) with complete analysis to verify that the HDE, HDI,
VDE and VDI values are inside the expected ranges.
Autocalibration
The autocalibration process provides an additional calibration of H FWHM and V FWHM values.
1. Open the Raw Data of the last QA phantom scan (The scan on the QA phantom must be performed on
the same day of the Autocalibration).
2. From the main toolbar select Primary reconstruction -> Auto Calibration and wait the end of the
process.
3. Once the process is finished run a new phantom analysis on the same scan.
4. Wait the end of the analysis process and verify that the H FWHM or the V FWHM values are inside the
expected ranges.
After the analysis is finished is recommended to perform a few check scan on the QA Phantom RawData in
order to have a visual check of the reconstructed axial images (in term of definition of borders). This control
should be done on the edges of the reconstructed area according to the followin table:
FOV Lower* Upper*
12” Between 5 and 6 Between 17.5 and 18.5
9” Between 2 and 3 Between 13 and 14
6” 1.5 10
Position of the Check Scan line referred to the RawData ruler (located on the right side of the scout).
NOTE (Multi-FOV devices):
After Autocalibration made on 12” FOV, could be possible that 9” QA Phantom test gives FWHM values out
of range or very close to the accepted limit. Repeat the Autocalibration on 9” FOV. At the end of calibration,
perform a QA Phantom scan and analysis on others FOV to check that new values are good for all fields.
Calibration Backup
At the end of the calibration a backup must be done in order to keep a copy of the modified files. Follow next
steps to perform the backup process.
1. From the main menu select File -> Settings Backup.
2. The “Import/Export path setup” window will appear. Select the “Change Path” button to select the
directory where you wish to backup the files. Use the “Add folder” button to create a new folder
inside the selected path.
3. Once the destination path is ready, select the “Go on” button to start the backup process.
4. At the end a message will appear informing that the Backup procedure has successfully completed.
5. Burn a CD with the contents of the directory in which you have just backed up the files.
NOTE: The CD containing the calibration files should be labeled with the serial number of the machine and
the date of the calibration and shall be kept in a safe place. Remember that these data are essential in case
of a computer failure or crash.

Daily Check
Following the list of tests performed by Daily Check procedure:
Test N. 1: Beam Limiter begins the rotation to Reset position (only multi-FOV devices)
Test N. 2: The Arm moves to 0 degrees
Test N. 3: Wait for the end of the rotation of the Beam Limiter (only multi-FOV devices)
Test N. 4: The Arm moves to 90 degrees (checks the required time)
Test N. 5: Sets the Large FOV of IB and checks his power voltages
Test N. 6: Switch on and checks the CCD camera
Test N. 7: Sets Large focal spot on the filament board
Test N. 8: Setup of the Xray system. Checks feedback signals from Inverter and Filament board. Check
connection Grabber – CCD camera.
Test N. 9: First of all, perform 10 acquisitions to fine adjust the diaphragm (as during Blank procedure) with
the Large FOV. Following table lists the expected gray level values (range ±5%):
Standard XRay tube
Detector
<03/06 >03/06
12” 2380 2800
9” 2600 2700
CSA
shell
XRay tube
Detector
<03/06 >03/06
12” 2400 2700
9” 2500 2600
After, another set of 5 acquisition is performed. The software calculates the mean gray level of
every image. At the end, the mean of the 5 means (MEAN) and the difference between the
maximum and minimum single mean are calculated (SD). For the MEAN treference values are
the same of previous table and for SD the maximum value is the 5% of the MEAN.
Test N. 10: Switch the Xray system off.
Test N. 11: Switch the CCD camera off.
Test N. 12: The Arm moves to 0 degrees

E-mail for Service: info@iviscans.com